The HCP8000 series current probes are wide band width DC/AC active current probes, featuring high bandwidth, fast and accurate capture the current wave, accuracy up to 1% and low circuit insertion loss. This probe can be used with any oscilloscope having a high-impedance BNC input.
Model | HCP8070 | ||
Bandwidth (-3dB) | DC-30MHz (Figure7) | ||
Rise time | ≤11.6ns | ||
Continuous maximum input range | 70Arms (Figure8) | ||
Max peak current value | 100Apk | ||
Range | 10A | 2X | |
70A | 20X | ||
Overload | 10A | ≥10APk | |
70A | ≥100APk | ||
Current transfer ratio | 10A | 0.5V/A | |
70A | 0.05V/A | ||
measurable current | 10A | 2mA | |
70A | 20mA | ||
Amplitude accuracy (DC,45-66Hz ) | 10A | ±1%±2mA | |
70A | ±1%±20mA | ||
Input impedance | Reference (Figure9) | ||
Delay time | Probe | 23ns | |
BNC (1m) | 5ns | ||
Terminal load | ≥100kΩ | ||
Power supply | DC 12V/1A(Standard Adaptor) | ||
Voltage of insulated wire | 300V CAT I | ||
Safety compliance | EN61010-1: 2010+A1:2019 EN 61010-2-032:2019 | ||
EMC standard | EN61326-1:2013 EN61000-3-2:2014 EN61000-3-3:2013 |