Model  | DP6070A  | DP6150A  | DP6350A  | DP6700A  | ||||||
Bandwidth(-3dB)  | 100MHz  | |||||||||
Rise time  | ≤3.5ns  | |||||||||
Accuracy  | ±2%  | |||||||||
Range selection (Attenuation rate)  | 10X/100X  | 50X/500X  | 100X/1000X  | 100X/1000X  | ||||||
Maximum differential voltage(DC + Peak AC)  | 10X  | ±70V  | 50X  | ±150V  | 100X  | ±350V  | 100X  | ±700V  | ||
100X  | ±700V  | 500X  | ±1500V  | 1000X  | ±3500V  | 1000X  | ±7000V  | |||
Common mode voltage (Vp) (DC + Peak AC)  | ±700V  | ±1500V  | ±3500V  | ±7000V  | ||||||
Maximum differential mode VS frequency curve  | Reference Figure 1  | Figure 2  | Figure 3  | Figure 4  | ||||||
Maximum input voltage-to-earth (Vrms)  | 450V CATII 600V CATI*  | 600V CATIII 1000V CATII  | 600V CATIII 1000V CATII  | 1000V CATIII 2300V CATI*  | ||||||
Input impedance  | Single-ended to ground  | 2.5MΩ  | 5MΩ  | 5MΩ  | 20MΩ  | |||||
Between inputs  | 5MΩ  | 10MΩ  | 10MΩ  | 40MΩ  | ||||||
Input capacitance  | Single-ended to ground  | <4pF  | <4pF  | <4pF  | <5pF  | |||||
Between inputs  | <2pF  | <2pF  | <2pF  | <2.5pF  | ||||||
CMRR  | DC  | >80dB  | >80dB  | >80dB  | >80dB  | |||||
100kHz  | >60dB  | >60dB  | >60dB  | >60dB  | ||||||
1MHz  | >50dB  | >50dB  | >50dB  | >50dB  | ||||||
Noise(Vrms)  | 10X  | <25mV  | 50X  | <50mV  | 100X  | <100mV  | 100X  | <220mV  | ||
100X  | <120mV  | 500X  | <300mV  | 1000X  | <600mV  | 1000X  | <1.2V  | |||
Differential overvoltage detection level  | 10X  | ≥70V  | 50X  | ≥150V  | 100X  | ≥350V  | 100X  | ≥700V  | ||
100X  | ≥700V  | 500X  | ≥1500V  | 1000X  | ≥3500V  | 1000X  | ≥7000V  | |||
Propagation time  | Probe  | 10X  | 8.7ns  | 50X  | 8.5ns  | 100X  | 8.2ns  | 100X  | 8.9ns  | |
100X  | 8.1ns  | 500X  | 7.5ns  | 1000X  | 7.4ns  | 1000X  | 6.6ns  | |||
BNC Line(1m)  | About 5ns  | |||||||||
Bandwidth limit filters (5MHz)  | ≥-3dB@5MHz  | |||||||||
Overload indicator (red light)  | Yes  | |||||||||
Overload alarm  | Yes(Can shut up manually)  | |||||||||
Automatic save  | Yes  | |||||||||
Offset setting function  | Yes( Set in test mode)  | |||||||||
Terminate load  | ≥100kΩ  | |||||||||
Power supply  | USB 5V/1A adapter  | |||||||||
Safety standard  | IEC/EN 61010-031:2015 + AMD1:2018  | |||||||||
EMC standard  | EN61326-1:2013 EN61000-3-2:2006+A1:2009+A2:2009 EN61000-3-3:2013  | |||||||||
*CAT I per IEC/EN 61010-031/A1:2008. No Rated Measurements Category per IEC/EN 61010-031:2015 + AMD1:2018.